Corpse Wiki

one) Regeneration fuel: a mixture of hydrogen and inert fuel (hydrogen accounts for five% -ten%)

2) Quantity: 40L gas cylinder 10Mpa

3) Regenerative gas strain: .05Mpa (the gasoline intake per fifty percent hour in the condition of .05Mpa is that the cylinder pressure drops by 1.5Mpa. If it exceeds one.5Mpa, the tension of the tension regulator’s timetable will be reduced, if it is a lot less than 1.5Mpa Improve the pressure of the strain regulator routine)

4) Process: 3 several hours heating-three hours heating and ventilation-three hours vacuuming-one hour pumping-10 hrs cooling

five) Measures:

① Join the regeneration fuel (examine the pipeline for leaks immediately after connection: following connecting, open up the major valve of the strain decreasing valve to the most, open the auxiliary meter to .1Mpa, convert off the major valve, innovative technology glovebox hold out 15 minutes, if the pressure of the auxiliary meter remains at .1 Mpa continues to be unchanged, indicating no air leakage)

②Set the strain of the auxiliary gauge of the tension decreasing valve to .1Mpa (the tension throughout air flow is about .06-.08MPa)

③ Close the cycle and analyzer

④Turn on the vacuum pump

⑤ Start regeneration

⑥ The new method ends at 20h (clean ahead of cleaning, drinking water and oxygen index <200ppm).

6) Note: Do not cut off the power and gas during the regeneration process (especially the first 6 hours)! Some pressure reducing valves are not allowed, and the pressure needs to be adjusted appropriately to ensure continuous ventilation for 3 hours!

7) Observation: whether the regeneration gas flowmeter shows about 20L / min, otherwise adjust the pressure of the pressure reducing valve.